논문 국내 국내전문학술지(KCI급) 분류식 습도 발생 장치 개발 및 라디오존데 습도센서 저온 효과 보정에 활용 연구

  • 학술지 구분 국내전문학술지(KCI급)
  • 게재년월 2021-09
  • 저자명 장은정, 이영석, 최병일, 최윤석, 이상욱
  • 학술지명 Journal of Sensor Science and Technology
  • 발행처명 한국센서학회
  • 발행국가 국내
  • 논문언어 한국어
  • 전체저자수 5

논문 초록 (Abstract)

Humidity is an important physical quantity that is closely related with the quality of everyday life as well as the quality control of products in various industries. Here, we have developed a divided-flow type humidity generator of which humidity generation is faster than the saturator-based humidity generator in ppm level. The operation principle of the divided-flow humidity generator is first introduced. Then, the performance of the divided-flow humidity generator is verified by testing the radiosonde humidity sensor at low temperature. As a result, the humidity generated from the divided-flow humidity generator is consistent with the saturator-based precision humidity generator within 1.6% relative humidity in the range from 10% to 40% at -45oC. It is also found that the radiosonde humidity sensor shows measurement errors by 3% - 5% at -45oC when it is only calibrated at room temperature. The response times of radiosonde humidity sensor using the divided-flow humidity generator are between about 2 and 9 minutes, whereas those by the saturatorbased humidity generator are about 20 minutes. In this regard, the divided-flow humidity generator has a merit in terms of fast humidity changes for the calibration of radiosonde humidity sensors at low-temperatures.